SCPI Command Reference
Variable Attenuator 135
Product-Specific Commands—Description
:OUTPut[1..n]:REFerence?
Description This query returns either the current or the
minimum/maximum output power reference
value.
At *RST, the value that will be set depends on the
instrument you have.
Syntax :OUTPut[1..n]:REFerence?[<wsp>MAXimum|
MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Reference>