IEEE 488.2 and Specific Commands
Integrated Qualification System 287
IEEE 488.2 Required Commands
*ESE
Description The *ESE command sets the Standard Event
Status Enable Register bits, as defined in the
table below. This register contains a mask value
for the bits to be enabled in the Standard Event
Status Register.
Syntax *ESE<wsp><RegisterValue>
Parameter(s) RegisterValue:
The program data syntax for <RegisterValue> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <RegisterValue>, expressed in base 2,
represents the bit values of the Standard Event
Status Enable Register.
CME
MSB
PON DDE N.U OPC
EXE QYE
LSB
Standard Event Status Enable Register