Exfo RTU 310-310G IP Services Test Head Manual do Utilizador Página 31

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 30
Getting Started
Ethernet Test Module 19
Connecting the Power
To connect the RTU-310/310G to a DC power source:
1. Remove the two Phillips flat screws, and remove the ground lug on the
back panel of the RTU-310/310G unit.
2. Use a #6 AWG wire (not supplied) insert the wire into the lug and
crimp it.
3. Use the two Phillips flat screws to attach the ground lug and wire
assembly to the rear panel of the RTU-310/310G unit.
4. Connect the other end to the ground distribution network.
WARNING
The RTU-310/310G DC version is intended to be grounded. Ensure
that the unit is connected to earth ground during normal use.
Vista de página 30