Exfo FTB-500 (Window XP) Manual do Utilizador Página 399

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 481
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 398
IEEE 488.2 and Specific Command Reference
FTB-500 391
IEEE 488.2 Required Commands
*OPC
Description The *OPC command makes synchronization
between the instrument and an external
controller possible. The *OPC command causes
the instrument to set bit 0 (Operation Complete)
in the Standard Event Status Register to the TRUE
(logic 1) state when the instrument completes all
pending operations. Detection of the Operation
Complete message can be accomplished by
continuous polling of the Standard Event Status
Register using the *ESR? common query
command. However, using a service request
eliminates the need to poll the Standard Event
Status Register thereby freeing the controller to
do other useful work.
Syntax *OPC
Parameter(s) None
See Also *OPC?
*WAI
Vista de página 398
1 2 ... 394 395 396 397 398 399 400 401 402 403 404 ... 480 481

Comentários a estes Manuais

Sem comentários