Exfo FTB-500 (Window XP) Manual do Utilizador Página 265

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 264
Maintenance
FTB-500 255
Recalibrating the Batteries
2. Connect your unit to a power outlet using the provided AC
adapter/charger or power supply.
3. Tap the Start Calibration button (the Stop Calibration button will
become available).
When a calibration step is underway, the on-screen LED preceding the
step name appears in yellow. When a step is complete, the LED turns
to green.
Once calibration is complete, the Start Calibration button becomes
available again.
Vista de página 264
1 2 ... 260 261 262 263 264 265 266 267 268 269 270 ... 404 405

Comentários a estes Manuais

Sem comentários