Exfo IQS-12002B Calibration System Manual do Utilizador Página 81

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 80
Calibration Methods
Calibration System 77
FTB-7000D and FTB-7000B-B OTDR Modules
Dead Zone Calibration
The dead zone depends on the intensity of the reflectance. Compliance to
the specification must be done for a reflectance of –35 dB in multimode
and –45 dB in singlemode.
Note: There is no uncertainty value associated with this calibration.
Attenuation Dead Zone
The attenuation dead zone is the distance between markers A and B
on the OTDR trace presented below.
Marker A is positioned at the rising edge of the reflective event.
Marker B is positioned within the falling edge, 0.5 dB above the
backscatter level.
Attenuation
dead zone
Vista de página 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 107 108

Comentários a estes Manuais

Sem comentários