Exfo FLS-2600B Tunable Laser Source Manual do Utilizador Página 143

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 142
Remote Control
Tunable Laser Source 135
SCPI Management Errors (System Errors)
SCPI Management Errors (System Errors)
Error
Number
Description Probable Cause
–100 Command error The SCPI Manager does not
recognize the command, probably
due to invalid module address.
104 Data-type error A data parameter is not the
expected data type.
108 Parameter not allowed Too many parameters were
detected for the specified
command.
109 Missing parameter A required parameter was not
detected.
113 Undefined header The SCPI Manager does not
recognize the command, probably
due to incorrect grammar.
130 Suffix error An error occurred while parsing a
suffix.
131 Invalid suffix The suffix does not follow the
appropriate syntax or is
inappropriate for this device.
138 Suffix not allowed A suffix was encountered after a
numeric element that does not
allow suffixes.
200 Execution error An error occurred while executing
the command.
222 Parameter out of range The Tunable Laser Source has
received a data parameter outside
the valid range.
Vista de página 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 156 157

Comentários a estes Manuais

Sem comentários