Exfo FTB-3930 MultiTest Manual do Utilizador Página 48

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 47
Performing Automated IL/ORL/Length Measurements (FasTesT)
42 FTB-3930
Referencing Units for FasTesT
4. Press Take Reference. After a few seconds, the unit displays new
reference values for each wavelength. If values are not acceptable, try
to clean connectors and repeat this step.
5. Disconnect the reference patchcord from the power meter adapter
only and connect it to the fiber under test.
6. Repeat the procedure with the second unit.
IMPORTANT
You can turn off the unit without losing the reference.
If you disconnect the patchcord from the FASTEST port, you must
take a new reference.
Vista de página 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 100 101

Comentários a estes Manuais

Sem comentários