Controlling the Source Remotely
CD/PMD Analyzer Source 27
Communication Parameters
Communication Parameters
The communication parameters are used to set the communication port.
Note: EOS means “End of String.” EOI means “End or Identify.”
For GPIB Communication
Terminate Read on EOS Yes
Set EOI with EOS on Writes Yes
Type of compare on EOS 8 bits
EOS byte 0Ah
Send EOI at end of Writes Yes
GPIB primary address See Setting the GPIB Address on
page 24
GPIB secondary address None
For RS-232 Communication
EOS bytes 0Ah
Baud rate 1200/2400/4800/9600/ 19200 bps
Parity None
Data bits 8 bits
Stop bits 1 bit
Flow control Software (Xon/Xoff) or None
Activation See Setting the Remote Command
Mode on page 23