Remote Control Commands
84 FLS-5834A
Product-Specific Commands-Description
SYSTem:ERRor?
Description This command returns the next error in the error queue. When
an error is generated, an error number is sent to the error
queue. The error queue is accessed with the SYST:ERR? query.
Syntax SYST:ERR?
Response See SCPI-Based Errors on page 85.
SYSTem:VERSion?
Description This query reads the CD/PMD Analyzer Source identification
string.
Syntax SYST:VERS?
Response “EXFO E.-O. Eng. FLS-5834A vx.xx”xxxxxxxx xxxxxxx, where
xxxxxxxx xxxxxxx is the serial number and vx.xx is the current
product version.
Note This query returns the same response as the *IDN? query.