User GuideOptical Spectrum Analyzer for FTB-500FTB-5240S/5240BP

Página 2

Introducing the FTB-5240S/5240BP Optical Spectrum Analyzer2 FTB-5240S/5240BPModelsModelsThe Optical Spectrum Analyzer comes in different models:³ FTB-

Página 3 - Contents

Setting Up the Instrument in WDM Mode92 FTB-5240S/5240BPSetting Up WDM Analysis Parameters4. In the Save As window, enter a file name and press Save.

Página 4

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 93Setting Up WDM Analysis ParametersTo import a test setup:1. From the Main Menu, press

Página 5

Setting Up the Instrument in WDM Mode94 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Favorites tab.3. Press Import to import an ana

Página 6 - Certification Information

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 95Setting Up WDM Analysis Parameters5. If you want to apply the above settings to the a

Página 7 - DECLARATION OF CONFORMITY

Setting Up the Instrument in WDM Mode96 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo delete a test configuration:1. From the Main Menu, press

Página 8

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 97Setting Up WDM Analysis Parameters2. Select the Favorites tab.

Página 9 - Spectrum Analyzer

Setting Up the Instrument in WDM Mode98 FTB-5240S/5240BPSetting Up WDM Analysis Parameters3. To delete a configuration file from the favorites list, s

Página 10

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 99Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing y

Página 11 - Typical Applications

Setting Up the Instrument in WDM Mode100 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main wi

Página 12

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 101Setting Up Acquisition Parameters³ If you are performing an Averaging or In-Band typ

Página 13 - 2 Safety Information

Introducing the FTB-5240S/5240BP Optical Spectrum AnalyzerOptical Spectrum Analyzer 3Typical ApplicationsTypical ApplicationsYou can use your Optical

Página 15 - Optical Spectrum Analyzer

Optical Spectrum Analyzer 1036 Setting Up the Instrument in Drift ModeBefore performing a spectral analysis in the Drift mode, you must set up your OS

Página 16 - 8 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode104 FTB-5240S/5240BP³ The easiest way to set up the instrument, especially when the operator does not know in a

Página 17 - FTB-500 left panel

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 105Defining PreferencesDefining PreferencesThe preferences window allows you to set g

Página 18 - 10 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode106 FTB-5240S/5240BPDefining Preferences2. Select the General tab.3. Define the general parameters as needed.4.

Página 19 - Optical Spectrum Analyzer 11

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 107Defining PreferencesTo enter link and location information:1. From the Main Menu,

Página 20 - Application

Setting Up the Instrument in Drift Mode108 FTB-5240S/5240BPDefining Preferences3. Under System and link information, define the following parameters a

Página 21

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 109Defining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Página 22 - Exiting the Application

Setting Up the Instrument in Drift Mode110 FTB-5240S/5240BPDefining Preferences2. Select the Comments tab. 3. Enter your comments for the current trac

Página 23 - Spectrum Analyzer for a Test

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 111Defining PreferencesDefining Display ParametersThe application allows you to set d

Página 24 - 16 FTB-5240S/5240BP

Introducing the FTB-5240S/5240BP Optical Spectrum Analyzer4 FTB-5240S/5240BPConventionsConventionsBefore using the product described in this manual, y

Página 25

Setting Up the Instrument in Drift Mode112 FTB-5240S/5240BPDefining Preferences2. Select the Display tab.

Página 26 - Selecting a Test Mode

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 113Defining Preferences3. Select the spectral unit you want to work with. With the na

Página 27 - To select a test mode:

Setting Up the Instrument in Drift Mode114 FTB-5240S/5240BPDefining Preferences4. Select the label that will appear on the peaks in the graph.

Página 28 - 20 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 115Defining PreferencesNote: The channel name and channel number cannot be shown at t

Página 29 - Nulling Electrical Offsets

Setting Up the Instrument in Drift Mode116 FTB-5240S/5240BPDefining Preferences5. Select Show to enable horizontal markers in the Markers tab. If Hide

Página 30 - 3. Press Nulling

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 117Defining Preferences6. Select the background color scheme for Graph as desired. 7.

Página 31

Setting Up the Instrument in Drift Mode118 FTB-5240S/5240BPDefining PreferencesCustomizing Drift Results TableIt is possible to select which results y

Página 32

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 119Defining Preferences2. Select the Drift Results tab. 3. Select which parameters yo

Página 33 - Performing User Calibration

Setting Up the Instrument in Drift Mode120 FTB-5240S/5240BPDefining PreferencesCustomizing Drift File NameThe application shall provide a way to defin

Página 34 - 26 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 121Defining Preferences2. Select the File Name tab. 3. Select which parameters you wa

Página 35

Optical Spectrum Analyzer 52 Safety InformationYour instrument is a Class 1 laser product in compliance with standards IEC 60825-1:2007 and 21 CFR 104

Página 36 - 28 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode122 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the field values will

Página 37

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 123Setting Up Drift Analysis ParametersSetting Up Drift Analysis ParametersThis secti

Página 38 - 30 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode124 FTB-5240S/5240BPSetting Up Drift Analysis ParametersDefining General SettingsThe general analysis parameter

Página 39

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 125Setting Up Drift Analysis Parameters2. Select the General tab.

Página 40 - 32 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode126 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. Under Default channel settings, define the following

Página 41

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 127Setting Up Drift Analysis Parametersand be considered as one multi-peak signal. Wi

Página 42 - To save a user calibration:

Setting Up the Instrument in Drift Mode128 FTB-5240S/5240BPSetting Up Drift Analysis Parameters³ Noise for OSNR: indicates which calculation method to

Página 43

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 129Setting Up Drift Analysis Parameters³ Peak detection level (dBm): indicates the mi

Página 44 - 36 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode130 FTB-5240S/5240BPSetting Up Drift Analysis Parameters³ Power offset (dB): indicates the offset value applied

Página 45 - WDM Mode

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 131Setting Up Drift Analysis ParametersYou can set your pass/fail threshold limits in

Página 47

Setting Up the Instrument in Drift Mode132 FTB-5240S/5240BPSetting Up Drift Analysis ParametersDefining Channel ThresholdsThe thresholds will be appli

Página 48 - 40 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 133Setting Up Drift Analysis ParametersWhen thresholds are globally disabled, the res

Página 49

Setting Up the Instrument in Drift Mode134 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Thresholds tab.

Página 50

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 135Setting Up Drift Analysis Parameters3. Select the Activate all thresholds option,

Página 51

Setting Up the Instrument in Drift Mode136 FTB-5240S/5240BPSetting Up Drift Analysis Parameters4. Enter values in the boxes as explained below:³ Wavel

Página 52 - 44 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 137Setting Up Drift Analysis ParametersManaging ChannelsTesting DWDM systems involves

Página 53

Setting Up the Instrument in Drift Mode138 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab. 3. By default, the channel

Página 54

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 139Setting Up Drift Analysis Parameters³ Stop range (nm or Thz): indicates the ending

Página 55

Setting Up the Instrument in Drift Mode140 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Press OK to return to the Channels window, which now

Página 56 - 48 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 141Setting Up Drift Analysis ParametersTo edit the parameters of a specific channel:1

Página 57

Optical Spectrum Analyzer 73 Getting Started with Your Optical Spectrum AnalyzerInserting and Removing Test ModulesTo insert a module into the FTB-500

Página 58 - 50 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode142 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab.

Página 59

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 143Setting Up Drift Analysis Parameters3. Select the channel or channels to be modifi

Página 60 - Customizing WDM Results Table

Setting Up the Instrument in Drift Mode144 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Modify the settings as needed. For more information

Página 61 - Optical Spectrum Analyzer 53

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 145Setting Up Drift Analysis ParametersTo add current peaks:Note: You can add current

Página 62 - 54 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode146 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab. 3. Press Import from Trace.

Página 63 - Customizing WDM File Name

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 147Setting Up Drift Analysis ParametersManaging FavoritesFavorites are configurations

Página 64

Setting Up the Instrument in Drift Mode148 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Favorites tab.

Página 65

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 149Setting Up Drift Analysis Parameters3. To apply the settings from a favorite file

Página 66 - 58 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode150 FTB-5240S/5240BPSetting Up Drift Analysis ParametersTo save a test configuration:1. From the Main Menu, pre

Página 67

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 151Setting Up Drift Analysis Parameters2. Select the Favorites tab.

Página 68

Getting Started with Your Optical Spectrum Analyzer8 FTB-5240S/5240BPInserting and Removing Test Modules3. Take the module and place it so that the co

Página 69 - Optical Spectrum Analyzer 61

Setting Up the Instrument in Drift Mode152 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. To save an analysis setup to a file, press Save As.

Página 70 - 62 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 153Setting Up Drift Analysis Parameters4. In the Save As window, enter a file name an

Página 71 - Optical Spectrum Analyzer 63

Setting Up the Instrument in Drift Mode154 FTB-5240S/5240BPSetting Up Drift Analysis ParametersTo import a test setup:1. From the Main Menu, press Ana

Página 72

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 155Setting Up Drift Analysis Parameters2. Select the Favorites tab.3. Press Import to

Página 73 - Optical Spectrum Analyzer 65

Setting Up the Instrument in Drift Mode156 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Press OK to load the configuration and close the win

Página 74 - 66 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 157Setting Up Drift Analysis Parameters2. Select the Favorites tab.

Página 75 - Defining Global Thresholds

Setting Up the Instrument in Drift Mode158 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. To delete a configuration file from the favorites li

Página 76 - To define global thresholds:

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 159Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performin

Página 77 - Optical Spectrum Analyzer 69

Setting Up the Instrument in Drift Mode160 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main

Página 78 - 70 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 161Setting Up Acquisition Parameters3. Select the acquisition type. ³ If you are perf

Página 79 - Defining Default Thresholds

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 9Inserting and Removing Test Modules7. While applying slight pressure to

Página 80

Setting Up the Instrument in Drift Mode162 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these bands of the spe

Página 81 - Optical Spectrum Analyzer 73

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 163Setting Up Acquisition Parameters3. Select a sampling unit and enter a sampling co

Página 82

Setting Up the Instrument in Drift Mode164 FTB-5240S/5240BPSetting Up Acquisition Parameters7. Select the Keep historical traces in subfolder option t

Página 83 - Optical Spectrum Analyzer 75

Optical Spectrum Analyzer 1657 Setting Up the Instrument in DFB ModeBefore performing a spectral analysis in the DFB mode, you must set up your OSA mo

Página 84

Setting Up the Instrument in DFB Mode166 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you to set general info

Página 85 - Optical Spectrum Analyzer 77

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 167Defining Preferences2. Select the General tab.3. Define the general parameters as ne

Página 86 - 78 FTB-5240S/5240BP

Setting Up the Instrument in DFB Mode168 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Main Menu, press Prefe

Página 87 - Optical Spectrum Analyzer 79

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 169Defining Preferences3. Under System and link information, define the following param

Página 88 - 4. Press Edit Selection

Setting Up the Instrument in DFB Mode170 FTB-5240S/5240BPDefining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Página 89 - Optical Spectrum Analyzer 81

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 171Defining Preferences2. Select the Comments tab. 3. Enter your comments for the curre

Página 90 - 82 FTB-5240S/5240BP

Getting Started with Your Optical Spectrum Analyzer10 FTB-5240S/5240BPInserting and Removing Test ModulesTo remove a module from the FTB-500:1. Exit T

Página 91

Setting Up the Instrument in DFB Mode172 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you to set display sett

Página 92 - 84 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 173Defining Preferences2. Select the Display tab.

Página 93 - Optical Spectrum Analyzer 85

Setting Up the Instrument in DFB Mode174 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. With the nanometer (nm

Página 94

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 175Defining Preferences4. Select the background color scheme for Graph as desired. 5. P

Página 95

Setting Up the Instrument in DFB Mode176 FTB-5240S/5240BPDefining PreferencesCustomizing DFB File NameThe application shall provide a way to define th

Página 96 - 88 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 177Defining Preferences2. Select the File Name tab. 3. Select which parameters you want

Página 97 - Optical Spectrum Analyzer 89

Setting Up the Instrument in DFB Mode178 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the field values will ap

Página 98

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 179Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing

Página 99

Setting Up the Instrument in DFB Mode180 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main wi

Página 100 - 92 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 181Setting Up Acquisition Parameters³ If you are performing an Averaging type acquisiti

Página 101 - To import a test setup:

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 11Inserting and Removing Test Modules5. Hold the module by its sides or b

Página 103 - Optical Spectrum Analyzer 95

Optical Spectrum Analyzer 1838 Setting Up the Instrument in Spectral Transmittance ModeBefore performing a spectral analysis in the Spectral Transmitt

Página 104 - 96 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode184 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you

Página 105 - 2. Select the Favorites tab

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 185Defining Preferences2. Select the General tab.3. Define the gener

Página 106 - 98 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode186 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Mai

Página 107 - Optical Spectrum Analyzer 99

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 187Defining Preferences3. Under System and link information, define

Página 108 - 100 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode188 FTB-5240S/5240BPDefining PreferencesTo enter comments:1. From the Main Menu, press Prefere

Página 109 - Optical Spectrum Analyzer 101

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 189Defining Preferences2. Select the Comments tab. 3. Enter your com

Página 110

Setting Up the Instrument in Spectral Transmittance Mode190 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you

Página 111 - Drift Mode

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 191Defining Preferences2. Select the Display tab.

Página 112 - 104 FTB-5240S/5240BP

ii FTB-5240S/5240BPCopyright © 2009–2010 EXFO Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or

Página 113 - Defining Preferences

Getting Started with Your Optical Spectrum Analyzer12 FTB-5240S/5240BPStarting the Optical Spectrum Analyzer ApplicationStarting the Optical Spectrum

Página 114 - 106 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode192 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. Wit

Página 115

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 193Defining Preferences4. Select Show to enable horizontal markers i

Página 116 - IMPORTANT

Setting Up the Instrument in Spectral Transmittance Mode194 FTB-5240S/5240BPDefining Preferences5. Select the background color scheme for Graph as des

Página 117 - To enter comments:

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 195Defining PreferencesCustomizing Spectral Transmittance File NameT

Página 118 - 110 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode196 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameter

Página 119 - Defining Display Parameters

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 197Defining Preferences4. Press Up or Down to change the order in wh

Página 120 - 2. Select the Display tab

Setting Up the Instrument in Spectral Transmittance Mode198 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis ParametersSetting Up Spectral T

Página 121

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 199Setting Up Spectral Transmittance Analysis ParametersDefining ST

Página 122 - 114 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode200 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters2. Select the ST Anal

Página 123

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 201Setting Up Spectral Transmittance Analysis Parameters3. Under Glo

Página 124 - 116 FTB-5240S/5240BP

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 13Starting the Optical Spectrum Analyzer ApplicationThe main window (show

Página 125

Setting Up the Instrument in Spectral Transmittance Mode202 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters³ Channel distance (G

Página 126 - 118 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 203Setting Up Spectral Transmittance Analysis ParametersTo edit the

Página 127

Setting Up the Instrument in Spectral Transmittance Mode204 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters4. If you want to app

Página 128 - Customizing Drift File Name

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 205Setting Up Acquisition ParametersSetting Up Acquisition Parameter

Página 129 - Optical Spectrum Analyzer 121

Setting Up the Instrument in Spectral Transmittance Mode206 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:

Página 130 - 122 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 207Setting Up Acquisition Parameters³ If you are performing an Avera

Página 131 - Optical Spectrum Analyzer 123

Setting Up the Instrument in Spectral Transmittance Mode208 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these

Página 132

Optical Spectrum Analyzer 2099 Setting Up the Instrument in EDFA ModeBefore performing a spectral analysis in the EDFA mode, you must set up your OSA

Página 133 - 2. Select the General tab

Setting Up the Instrument in EDFA Mode210 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you to set general inf

Página 134 - 126 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 211Defining Preferences2. Select the General tab.3. Define the general parameters as n

Página 135 - Optical Spectrum Analyzer 127

Getting Started with Your Optical Spectrum Analyzer14 FTB-5240S/5240BPExiting the ApplicationStatus BarThe status bar, located at the bottom of the ma

Página 136 - 128 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode212 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Main Menu, press Pref

Página 137 - Optical Spectrum Analyzer 129

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 213Defining Preferences2. Select the Information tab.3. Under System and link informat

Página 138 - press Edit % button

Setting Up the Instrument in EDFA Mode214 FTB-5240S/5240BPDefining Preferences4. Under Location Information, define the following parameters as needed

Página 139 - Threshold Limit Definition

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 215Defining Preferences2. Select the Comments tab. 3. Enter your comments for the curr

Página 140 - Defining Channel Thresholds

Setting Up the Instrument in EDFA Mode216 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you to set display set

Página 141 - Optical Spectrum Analyzer 133

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 217Defining Preferences2. Select the Display tab.

Página 142 - 134 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode218 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. With the nanometer (n

Página 143 - Optical Spectrum Analyzer 135

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 219Defining Preferences4. Select the label that will appear on the peaks in the graph.

Página 144 - 136 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode220 FTB-5240S/5240BPDefining PreferencesNote: The channel name and channel number cannot be shown at the same ti

Página 145 - Managing Channels

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 221Defining Preferences5. Select Show to display the empty channels from the channel l

Página 146 - 138 FTB-5240S/5240BP

Optical Spectrum Analyzer 154 Preparing Your Optical Spectrum Analyzer for a TestCleaning and Connecting Optical FibersTo connect the fiber-optic cabl

Página 147 - Optical Spectrum Analyzer 139

Setting Up the Instrument in EDFA Mode222 FTB-5240S/5240BPDefining Preferences6. Select Show to enable horizontal markers in the Markers tab. If Hide

Página 148 - 140 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 223Defining Preferences7. Select the background color scheme for Graph as desired. 8.

Página 149 - Optical Spectrum Analyzer 141

Setting Up the Instrument in EDFA Mode224 FTB-5240S/5240BPDefining PreferencesCustomizing EDFA Results TableIt is possible to select which results you

Página 150 - 2. Select the Channels tab

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 225Defining Preferences2. Select the EDFA Results tab. 3. Select which parameters you

Página 151 - Optical Spectrum Analyzer 143

Setting Up the Instrument in EDFA Mode226 FTB-5240S/5240BPDefining Preferences³ Gain: indicates the gain (Output Signal Power - Input Signal Power) fo

Página 152 - 144 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 227Defining PreferencesCustomizing EDFA File NameThe application shall provide a way t

Página 153 - To add current peaks:

Setting Up the Instrument in EDFA Mode228 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameters you want to incl

Página 154 - 146 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 229Defining Preferences4. Press Up or Down to change the order in which the field valu

Página 155 - Managing Favorites

Setting Up the Instrument in EDFA Mode230 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersSetting Up EDFA Analysis ParametersThis section presents

Página 156

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 231Setting Up EDFA Analysis ParametersDefining General SettingsThe general analysis pa

Página 157 - Optical Spectrum Analyzer 149

Preparing Your Optical Spectrum Analyzer for a Test16 FTB-5240S/5240BPCleaning and Connecting Optical Fibers3. Carefully align the connector and port

Página 158 - To save a test configuration:

Setting Up the Instrument in EDFA Mode232 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters2. Select the General tab.

Página 159

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 233Setting Up EDFA Analysis Parameters3. Under Default channel settings, define the fo

Página 160 - 152 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode234 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters³ Channel width (GHz or nm): indicates the limit inside w

Página 161 - Optical Spectrum Analyzer 153

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 235Setting Up EDFA Analysis Parameters³ Peak detection level (dBm): indicates the mini

Página 162

Setting Up the Instrument in EDFA Mode236 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters³ Output wavelength offset (nm): indicates the offset val

Página 163 - Optical Spectrum Analyzer 155

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 237Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Página 164 - 156 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode238 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersManaging ChannelsTesting DWDM systems involves characteri

Página 165

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 239Setting Up EDFA Analysis Parameters2. Select the Channels tab. 3. By default, the c

Página 166 - 158 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode240 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters4. Enter values in the boxes as explained below:³ Start r

Página 167 - Optical Spectrum Analyzer 159

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 241Setting Up EDFA Analysis Parameters³ Starting Value: Sets the increment starting va

Página 168 - 160 FTB-5240S/5240BP

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 17Installing the EXFO Universal Interface (EUI)Installing the EXFO Univer

Página 169 - Optical Spectrum Analyzer 161

Setting Up the Instrument in EDFA Mode242 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters7. Press OK to save the changes and close the window, or

Página 170

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 243Setting Up EDFA Analysis Parameters2. Select the Channels tab.

Página 171 - Optical Spectrum Analyzer 163

Setting Up the Instrument in EDFA Mode244 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. Select the channel or channels to be modified in the c

Página 172 - 164 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 245Setting Up EDFA Analysis Parameters5. Modify the settings as needed. For more infor

Página 173 - DFB Mode

Setting Up the Instrument in EDFA Mode246 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters7. If you want to apply the above settings to the active

Página 174

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 247Setting Up EDFA Analysis ParametersTo add current peaks:Note: You can add current p

Página 175

Setting Up the Instrument in EDFA Mode248 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. Press Import from Trace. All peaks from the current tr

Página 176 - 168 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 249Setting Up EDFA Analysis Parameters4. If you want to apply the above settings to th

Página 177

Setting Up the Instrument in EDFA Mode250 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersManaging FavoritesFavorites are configurations files that

Página 178

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 251Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Página 179

Preparing Your Optical Spectrum Analyzer for a Test18 FTB-5240S/5240BPSelecting a Test ModeSelecting a Test ModeYour Optical Spectrum Analyzer gives y

Página 180

Setting Up the Instrument in EDFA Mode252 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To apply the settings from a favorite file to the curr

Página 181

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 253Setting Up EDFA Analysis Parameters4. If you want to apply the above settings to th

Página 182 - 174 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode254 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo save a test configuration:1. From the Main Menu, press

Página 183

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 255Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Página 184 - Customizing DFB File Name

Setting Up the Instrument in EDFA Mode256 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To save an analysis setup to a file, press Save As. Th

Página 185 - Optical Spectrum Analyzer 177

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 257Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Página 186 - 178 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode258 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo import a test setup:1. From the Main Menu, press Analy

Página 187 - Optical Spectrum Analyzer 179

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 259Setting Up EDFA Analysis Parameters2. Select the Favorites tab.3. Press Import to i

Página 188 - 180 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode260 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters4. From the Import window, select the file you want to im

Página 189 - Optical Spectrum Analyzer 181

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 261Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Página 190

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 19Selecting a Test ModeTo select a test mode:1. From the main menu, press

Página 191 - Spectral Transmittance Mode

Setting Up the Instrument in EDFA Mode262 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo delete a test configuration:1. From the Main Menu, pre

Página 192

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 263Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Página 193

Setting Up the Instrument in EDFA Mode264 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To delete a configuration file from the favorites list

Página 194 - 186 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 265Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing

Página 195

Setting Up the Instrument in EDFA Mode266 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main w

Página 196

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 267Setting Up Acquisition Parameters³ If you are performing an Averaging type acquisit

Página 197

Setting Up the Instrument in EDFA Mode268 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these bands of the spec

Página 198

Optical Spectrum Analyzer 26910 Testing DWDM SystemsOptical spectrum analysis is the measurement of optical power as a function of wavelength or frequ

Página 199

Testing DWDM Systems270 FTB-5240S/5240BPUsing the Discover Featurethan 0.001 nm or 0.001 GHz, then their width is reduce to the lower width. If the wi

Página 200 - 192 FTB-5240S/5240BP

Testing DWDM SystemsOptical Spectrum Analyzer 271Using the Discover FeatureTo start an automatic setup scan:From the Main Menu, press Discover.WDM Mod

Página 201

Preparing Your Optical Spectrum Analyzer for a Test20 FTB-5240S/5240BPSelecting a Test Mode2. Select the desired test mode.Once you select the mode, y

Página 202 - 194 FTB-5240S/5240BP

Testing DWDM Systems272 FTB-5240S/5240BPUsing the Discover FeatureNote: If you already have an active trace on screen that was modified, you will be p

Página 203

Testing DWDM SystemsOptical Spectrum Analyzer 273Starting a MeasurementStarting a MeasurementNote: Before starting a measurement you must select and c

Página 204

Testing DWDM Systems274 FTB-5240S/5240BPStarting a MeasurementDrift ModeDFB Mode

Página 205

Testing DWDM SystemsOptical Spectrum Analyzer 275Starting a MeasurementSpectral TransmittanceModeEDFA Mode

Página 206 - Parameters

Testing DWDM Systems276 FTB-5240S/5240BPStarting a MeasurementWhen an acquisition is undergoing in the module, the Start button changes to Stop. Press

Página 207 - Defining ST Analysis Settings

Optical Spectrum Analyzer 27711 Managing ResultsYour Optical Spectrum Analyzer allows you to manage results in the WDM, Drift, DFB, Spectral Transmitt

Página 208 - 200 FTB-5240S/5240BP

Managing Results278 FTB-5240S/5240BPManaging WDM Test ResultsManaging WDM Test ResultsThe application allows you to view and manage your WDM test resu

Página 209 - Optical Spectrum Analyzer 201

Managing ResultsOptical Spectrum Analyzer 279Managing WDM Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detail

Página 210

Managing Results280 FTB-5240S/5240BPManaging WDM Test ResultsYour application allows you to toggle the view of your WDM main window. You can change th

Página 211 - Optical Spectrum Analyzer 203

Managing ResultsOptical Spectrum Analyzer 281Managing WDM Test ResultsFor 100 % lower tabs view, press .The graph will display peak indicators for al

Página 212 - 204 FTB-5240S/5240BP

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 21Nulling Electrical OffsetsNulling Electrical OffsetsThe offset nulling

Página 213 - Optical Spectrum Analyzer 205

Managing Results282 FTB-5240S/5240BPManaging WDM Test ResultsThe graph will also display the noise level for a channel with a doted line under the sel

Página 214 - 206 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 283Managing WDM Test ResultsResults TabThe results table shows channel results for the active trace. The res

Página 215 - Optical Spectrum Analyzer 207

Managing Results284 FTB-5240S/5240BPManaging WDM Test Results³ Noise (dBm): indicates the level of the noise for the selected channel (according to th

Página 216

Managing ResultsOptical Spectrum Analyzer 285Managing WDM Test ResultsMarkers TabYou can use markers to perform manual measurements and verification d

Página 217 - EDFA Mode

Managing Results286 FTB-5240S/5240BPManaging WDM Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the marker by en

Página 218

Managing ResultsOptical Spectrum Analyzer 287Managing WDM Test ResultsYou can also move the markers directly on the Graph tab. Drag the marker to the

Página 219

Managing Results288 FTB-5240S/5240BPManaging WDM Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details about the acquisit

Página 220 - 212 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 289Managing WDM Test Results3. Select the General tab.4. Edit the general information as required.Note: If y

Página 221

Managing Results290 FTB-5240S/5240BPManaging WDM Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press Tr

Página 222 - 214 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 291Managing WDM Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Tr

Página 223

ContentsOptical Spectrum Analyzer iiiContentsCertification Information ...

Página 224

Preparing Your Optical Spectrum Analyzer for a Test22 FTB-5240S/5240BPNulling Electrical Offsets3. Press Nulling. You are notified that the nulling is

Página 225

Managing Results292 FTB-5240S/5240BPManaging WDM Test ResultsChannel Results TabWhen you select a row from the Results tab, the Channel Results tab wi

Página 226 - 218 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 293Managing WDM Test Results2. Select a row from the Results tab to view the channel results for the selecte

Página 227

Managing Results294 FTB-5240S/5240BPManaging WDM Test Results³ Bandwidth 3.00 dB: indicates bandwidth measured by taking the width of a signal at 50 %

Página 228 - 220 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 295Managing WDM Test Results³ Noise for OSNR: indicates which calculation method to use for OSNR value. Fixe

Página 229

Managing Results296 FTB-5240S/5240BPManaging WDM Test ResultsYou can include selected channels results in the global results. These channels will be i

Página 230 - 222 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 297Managing WDM Test Results2. Select a row from the Results tab to view the channel results for the selecte

Página 231

Managing Results298 FTB-5240S/5240BPManaging WDM Test Results3. Select the Include in calculation option.

Página 232 - 224 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 299Managing WDM Test ResultsTo re-analyze channel results:1. From the main window, select the Channel Result

Página 233 - Optical Spectrum Analyzer 225

Managing Results300 FTB-5240S/5240BPManaging WDM Test Results3. Edit the values as explained below:³ Channel number: indicates the number of the chann

Página 234 - 226 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 301Managing WDM Test Results³ Signal power calculation: indicates which calculation method to apply for sign

Página 235 - Customizing EDFA File Name

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 23Performing User CalibrationPerforming User CalibrationCalibrating your

Página 236

Managing Results302 FTB-5240S/5240BPManaging WDM Test Results4. Edit the thresholds as explained below:³ Wavelength/Frequency (nm/GHz): indicates the

Página 237

Managing ResultsOptical Spectrum Analyzer 303Managing WDM Test ResultsNote: If a channel is already included in the channel list, the application will

Página 238 - 230 FTB-5240S/5240BP

Managing Results304 FTB-5240S/5240BPManaging WDM Test ResultsGlobal Results TabThe application allows you to view the global results of the current me

Página 239

Managing ResultsOptical Spectrum Analyzer 305Managing WDM Test Results³ Total power in scan range: indicates the total power for the selected scan ran

Página 240

Managing Results306 FTB-5240S/5240BPManaging WDM Test ResultsChanging Active Trace Analysis Parameters and Re-analyzingThe global parameters modifies

Página 241 - Optical Spectrum Analyzer 233

Managing ResultsOptical Spectrum Analyzer 307Managing WDM Test Results2. Press Global Parameters. 3. Select the Analysis tab. 4. Enter values in the b

Página 242 - 234 FTB-5240S/5240BP

Managing Results308 FTB-5240S/5240BPManaging WDM Test Results³ Wavelength offset (nm): indicates the offset value applied on the wavelength. You can u

Página 243 - Optical Spectrum Analyzer 235

Managing ResultsOptical Spectrum Analyzer 309Managing WDM Test ResultsChanging Default Channel ParametersDefault channel settings will define how any

Página 244 - 236 FTB-5240S/5240BP

Managing Results310 FTB-5240S/5240BPManaging WDM Test Results2. Press Global Parameters. 3. Select the Default Channel tab.

Página 245 - Optical Spectrum Analyzer 237

Managing ResultsOptical Spectrum Analyzer 311Managing WDM Test Results4. Enter values in the boxes as explained below:³ Clear the Activate default cha

Página 246

Preparing Your Optical Spectrum Analyzer for a Test24 FTB-5240S/5240BPPerforming User CalibrationA high level sequence of steps to perform user calibr

Página 247

Managing Results312 FTB-5240S/5240BPManaging WDM Test Results³ Signal power calculation: indicates which calculation method to apply for signal power

Página 248

Managing ResultsOptical Spectrum Analyzer 313Managing WDM Test Results5. Press OK to save the changes, re-analyze data and close the window, or press

Página 249 - Optical Spectrum Analyzer 241

Managing Results314 FTB-5240S/5240BPManaging WDM Test ResultsTo change global threshold parameters:1. From the main window, select the Global Results

Página 250 - 242 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 315Managing WDM Test Results2. Press Global Parameters. 3. Select the Global Thresholds tab. 4. Select the A

Página 251

Managing Results316 FTB-5240S/5240BPManaging WDM Test Results³ Average OSNR (dB): indicates the sum of the entire OSNR of the peaks detected in the cu

Página 252 - 244 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 317Managing WDM Test ResultsChanging Default Thresholds ParametersThe default thresholds will be applied to

Página 253 - Optical Spectrum Analyzer 245

Managing Results318 FTB-5240S/5240BPManaging WDM Test Results2. Press Global Parameters. 3. Select the Default Thresholds tab.

Página 254 - 246 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 319Managing WDM Test Results4. Enter values in the boxes as explained below:³ Wavelength/Frequency (nm/GHz):

Página 255 - Optical Spectrum Analyzer 247

Managing Results320 FTB-5240S/5240BPManaging Drift Test ResultsManaging Drift Test ResultsThe application allows you to view and manage your drift tes

Página 256 - 248 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 321Managing Drift Test ResultsYou can select a channel directly from the dashboard or from the Channel Histo

Página 257 - Optical Spectrum Analyzer 249

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 25Performing User CalibrationAn example of the calibration file format is

Página 258

Managing Results322 FTB-5240S/5240BPManaging Drift Test Results³ The dashboard shows a global status (all channels) for each parameter. This global st

Página 259

Managing ResultsOptical Spectrum Analyzer 323Managing Drift Test Results³ The dashboard displays a channel status (all parameters) for a given channel

Página 260 - 252 FTB-5240S/5240BP

Managing Results324 FTB-5240S/5240BPManaging Drift Test ResultsChannel Graph TabThe Channel Graph tab displays three different graphs for the selected

Página 261 - Optical Spectrum Analyzer 253

Managing ResultsOptical Spectrum Analyzer 325Managing Drift Test ResultsChannel History TabThe Channel History table shows channel results for the act

Página 262

Managing Results326 FTB-5240S/5240BPManaging Drift Test ResultsTo view channel history results:From the main window, select the Channel History tab.Re

Página 263

Managing ResultsOptical Spectrum Analyzer 327Managing Drift Test ResultsA small red marker ( ) will point down at the peak in the WDM Graph tab when y

Página 264 - 256 FTB-5240S/5240BP

Managing Results328 FTB-5240S/5240BPManaging Drift Test ResultsWDM Graph TabThe WDM Graph tab allows you to view the spectrum of the active trace for

Página 265 - Optical Spectrum Analyzer 257

Managing ResultsOptical Spectrum Analyzer 329Managing Drift Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for deta

Página 266

Managing Results330 FTB-5240S/5240BPManaging Drift Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly

Página 267

Managing ResultsOptical Spectrum Analyzer 331Managing Drift Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the m

Página 268 - Favorites window

Preparing Your Optical Spectrum Analyzer for a Test26 FTB-5240S/5240BPPerforming User Calibration2. Select the Calibration tab. Note: You cannot edit

Página 269 - Optical Spectrum Analyzer 261

Managing Results332 FTB-5240S/5240BPManaging Drift Test ResultsYou can also move the markers directly on the WDM Graph tab. Drag the marker to the des

Página 270 - 262 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 333Managing Drift Test ResultsTo edit general information:1. From the main window, select the Trace Info. ta

Página 271

Managing Results334 FTB-5240S/5240BPManaging Drift Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press

Página 272 - 264 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 335Managing Drift Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press

Página 273 - Optical Spectrum Analyzer 265

Managing Results336 FTB-5240S/5240BPManaging Drift Test ResultsChannel Results TabWhen you select a channel in the Channel History tab, the Channel Re

Página 274 - 266 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 337Managing Drift Test Results2. Select a channel from the Channel History tab to view the channel results f

Página 275 - Optical Spectrum Analyzer 267

Managing Results338 FTB-5240S/5240BPManaging Drift Test Results³ Bandwidth x dB: indicates bandwidth measured by taking the width of a signal at x dB

Página 276

Managing ResultsOptical Spectrum Analyzer 339Managing Drift Test ResultsIn-Band narrow filter (InB nf): The In-Band narrow filter method uses the In-B

Página 277 - 10 Testing DWDM Systems

Managing Results340 FTB-5240S/5240BPManaging DFB Test ResultsManaging DFB Test ResultsThe application allows you to view and manage your DFB test resu

Página 278 - Using the Discover Feature

Managing ResultsOptical Spectrum Analyzer 341Managing DFB Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detail

Página 279 - Testing DWDM Systems

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 27Performing User Calibration3. If a user calibration file exists, press

Página 280 - 272 FTB-5240S/5240BP

Managing Results342 FTB-5240S/5240BPManaging DFB Test ResultsResults TabYou can view the analysis of the DFB laser source from the Results tab.To v ie

Página 281 - Starting a Measurement

Managing ResultsOptical Spectrum Analyzer 343Managing DFB Test Results³ Right stopband: indicates the spectral position difference between the main mo

Página 282

Managing Results344 FTB-5240S/5240BPManaging DFB Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly o

Página 283

Managing ResultsOptical Spectrum Analyzer 345Managing DFB Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the mar

Página 284 - 276 FTB-5240S/5240BP

Managing Results346 FTB-5240S/5240BPManaging DFB Test ResultsNote: After using the zoom tools in the DFB Graph tab, you can only move the markers agai

Página 285 - 11 Managing Results

Managing ResultsOptical Spectrum Analyzer 347Managing DFB Test Results3. Select the General tab.4. Edit the general information as required.Note: If y

Página 286 - Managing WDM Test Results

Managing Results348 FTB-5240S/5240BPManaging DFB Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press Tr

Página 287

Managing ResultsOptical Spectrum Analyzer 349Managing DFB Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Tr

Página 288 - Managing Results

Managing Results350 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsManaging Spectral Transmittance Test ResultsThe application allows you

Página 289

Managing ResultsOptical Spectrum Analyzer 351Managing Spectral Transmittance Test ResultsWhen the acquisition is taken (see Starting a Measurement on

Página 290

Preparing Your Optical Spectrum Analyzer for a Test28 FTB-5240S/5240BPPerforming User CalibrationThe calibration values will replace the Correction fa

Página 291 - Results Tab

Managing Results352 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsResults TabThe results table shows the spectral transmittance results

Página 292 - 284 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 353Managing Spectral Transmittance Test ResultsMarkers TabYou can use markers to perform manual measurements

Página 293 - Markers Tab

Managing Results354 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. S

Página 294

Managing ResultsOptical Spectrum Analyzer 355Managing Spectral Transmittance Test ResultsYou can also move the markers directly on the ST Graph tab. D

Página 295

Managing Results356 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details

Página 296 - Trace Info. Tab

Managing ResultsOptical Spectrum Analyzer 357Managing Spectral Transmittance Test Results3. Select the General tab.4. Edit the general information as

Página 297

Managing Results358 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTo edit trace information:1. From the main window, select the Trace In

Página 298 - 290 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 359Managing Spectral Transmittance Test ResultsTo edit comments:1. From the main window, select the Trace In

Página 299

Managing Results360 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsST Results TabThe ST Results tab will show complete information about

Página 300 - Channel Results Tab

Managing ResultsOptical Spectrum Analyzer 361Managing Spectral Transmittance Test Results³ Insertion loss max (dB): indicates the maximum difference b

Página 301 - Optical Spectrum Analyzer 293

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 29Performing User Calibration5. Press Write to Module to apply the modifi

Página 302

Managing Results362 FTB-5240S/5240BPManaging Spectral Transmittance Test Results³ Input wavelength offset (nm): indicates the offset value applied on

Página 303

Managing ResultsOptical Spectrum Analyzer 363Managing Spectral Transmittance Test ResultsTo edit ST analysis parameters:1. From the main window, selec

Página 304 - 296 FTB-5240S/5240BP

Managing Results364 FTB-5240S/5240BPManaging Spectral Transmittance Test Results2. Press Global Parameters.

Página 305

Managing ResultsOptical Spectrum Analyzer 365Managing Spectral Transmittance Test Results3. Edit the values as explained below:³ Channel definition (G

Página 306

Managing Results366 FTB-5240S/5240BPManaging Spectral Transmittance Test Results³ Channel range (GHz or nm): indicates the limit inside which the powe

Página 307 - 2. Press Channel Parameters

Managing ResultsOptical Spectrum Analyzer 367Managing Spectral Transmittance Test Results³ Output wavelength offset (nm): indicates the offset value a

Página 308

Managing Results368 FTB-5240S/5240BPManaging EDFA Test ResultsManaging EDFA Test ResultsThe application allows you to view and manage your EDFA test r

Página 309

Managing ResultsOptical Spectrum Analyzer 369Managing EDFA Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detai

Página 310

Managing Results370 FTB-5240S/5240BPManaging EDFA Test ResultsResults TabThe results table shows channel results for the active trace. The results for

Página 311

Managing ResultsOptical Spectrum Analyzer 371Managing EDFA Test Results³ Noise Figure: indicates the EDFA’s noise figure measured for the selected cha

Página 312 - Global Results Tab

Preparing Your Optical Spectrum Analyzer for a Test30 FTB-5240S/5240BPPerforming User Calibration6. To verify that the calibration changes are properl

Página 313 - Optical Spectrum Analyzer 305

Managing Results372 FTB-5240S/5240BPManaging EDFA Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly

Página 314 - Re-analyzing

Managing ResultsOptical Spectrum Analyzer 373Managing EDFA Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the ma

Página 315

Managing Results374 FTB-5240S/5240BPManaging EDFA Test ResultsYou can also move the markers directly on the EDFA Graph tab. Drag the marker to the des

Página 316 - 308 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 375Managing EDFA Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details about th

Página 317

Managing Results376 FTB-5240S/5240BPManaging EDFA Test Results3. Select the General tab.4. Edit the general information as required.Note: If you want

Página 318 - 2. Press Global Parameters

Managing ResultsOptical Spectrum Analyzer 377Managing EDFA Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2

Página 319

Managing Results378 FTB-5240S/5240BPManaging EDFA Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Trace Iden

Página 320 - 312 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 379Managing EDFA Test ResultsChannel Results TabWhen you select a row from the Results tab, the Channel Resu

Página 321

Managing Results380 FTB-5240S/5240BPManaging EDFA Test Results2. Select a row from the Results tab to view the channel results for the selected channe

Página 322

Managing ResultsOptical Spectrum Analyzer 381Managing EDFA Test Results³ Channel index: indicates the channel number. Channel 1 corresponds to the fir

Página 323

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 31Performing User CalibrationTo clear user correction factors:1. From the

Página 324 - 316 FTB-5240S/5240BP

Managing Results382 FTB-5240S/5240BPManaging EDFA Test ResultsYou can include selected channels results in the global results calculation. To include

Página 325

Managing ResultsOptical Spectrum Analyzer 383Managing EDFA Test Results

Página 326

Managing Results384 FTB-5240S/5240BPManaging EDFA Test Results3. Select the Include in calculation option.

Página 327 - Optical Spectrum Analyzer 319

Managing ResultsOptical Spectrum Analyzer 385Managing EDFA Test ResultsTo modify channel results parameters:1. From the main window, select the Channe

Página 328 - Managing Drift Test Results

Managing Results386 FTB-5240S/5240BPManaging EDFA Test Results

Página 329 - ³ Signal power

Managing ResultsOptical Spectrum Analyzer 387Managing EDFA Test Results3. Edit the values as explained below:³ Channel Center (nm or THz): indicates t

Página 330

Managing Results388 FTB-5240S/5240BPManaging EDFA Test ResultsGlobal Results TabThe application allows you to view the global results of the current m

Página 331

Managing ResultsOptical Spectrum Analyzer 389Managing EDFA Test Results³ Average gain: indicates the sum of the gain of all detected channels in the c

Página 332 - Channel Graph Tab

Managing Results390 FTB-5240S/5240BPManaging EDFA Test ResultsChanging Active Trace Analysis ParametersThe global parameters modifies the active trace

Página 333 - Channel History Tab

Managing ResultsOptical Spectrum Analyzer 391Managing EDFA Test Results3. Under Default channel settings, modify the following parameters as needed: ³

Página 334

Contentsiv FTB-5240S/5240BP9 Setting Up the Instrument in EDFA Mode ...209Defining Preferences ...

Página 335

Preparing Your Optical Spectrum Analyzer for a Test32 FTB-5240S/5240BPPerforming User Calibration2. Select the Calibration tab.

Página 336 - WDM Graph Tab

Managing Results392 FTB-5240S/5240BPManaging EDFA Test Results4. Under Global analysis parameters, modify the following parameters as needed:³ Peak de

Página 337 - Optical Spectrum Analyzer 329

Managing ResultsOptical Spectrum Analyzer 393Managing EDFA Test Results³ Input power offset (dB): indicates the offset value applied on the input powe

Página 338

Managing Results394 FTB-5240S/5240BPManaging EDFA Test Results³ Output power offset (dB): indicates the offset value applied on the output power. You

Página 339 - Optical Spectrum Analyzer 331

Managing ResultsOptical Spectrum Analyzer 395Managing EDFA Test ResultsSwapping EDFA TracesThe Swap Trace feature allows you to swap EDFA input and ou

Página 340

Managing Results396 FTB-5240S/5240BPUsing Zoom ControlsUsing Zoom ControlsUse the zoom controls to change the scale of the trace display.You can zoom

Página 341

Managing ResultsOptical Spectrum Analyzer 397Using Zoom Controls³ You can zoom in or out on the center of the portion of the graph that is displayed b

Página 342 - 334 FTB-5240S/5240BP

Managing Results398 FTB-5240S/5240BPManaging Trace FilesManaging Trace FilesThe application allows you to manage the trace files for all test modes. Y

Página 343

Managing ResultsOptical Spectrum Analyzer 399Managing Trace Files2. Press New.

Página 344

Managing Results400 FTB-5240S/5240BPManaging Trace Files3. If you had already acquired (but not saved) a trace, a warning window appears, asking you i

Página 345 - Optical Spectrum Analyzer 337

Managing ResultsOptical Spectrum Analyzer 401Managing Trace FilesOpening Trace FilesOpening a trace file saves you time as you do not need to perform

Página 346

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 33Performing User Calibration3. Press Clear user correction factors, to l

Página 347 - Optical Spectrum Analyzer 339

Managing Results402 FTB-5240S/5240BPManaging Trace FilesOpening ST, EDFA and Legacy OSA trace files in WDM modeYour application allows you to open dif

Página 348 - Managing DFB Test Results

Managing ResultsOptical Spectrum Analyzer 403Managing Trace FilesIf the file contains two traces (Input and Output), then the application allows you t

Página 349 - Optical Spectrum Analyzer 341

Managing Results404 FTB-5240S/5240BPManaging Trace FilesOpening WDM trace file in DFB modeYour application allows you to open WDM file type in DFB mod

Página 350

Managing ResultsOptical Spectrum Analyzer 405Managing Trace FilesOpening WDM trace file in ST modeYour application allows you to open WDM file type in

Página 351 - Optical Spectrum Analyzer 343

Managing Results406 FTB-5240S/5240BPManaging Trace Files

Página 352

Managing ResultsOptical Spectrum Analyzer 407Managing Trace FilesOpening WDM trace file in EDFA modeYour application allows you to open WDM file type

Página 353 - Optical Spectrum Analyzer 345

Managing Results408 FTB-5240S/5240BPManaging Trace FilesAn example of WDM trace file imported in EDFA mode is shown below.

Página 354

Managing ResultsOptical Spectrum Analyzer 409Managing Trace FilesSaving Trace FilesThe application allows you to save the trace files for future refer

Página 355

Managing Results410 FTB-5240S/5240BPGenerating ReportsGenerating ReportsAfter performing any acquisition, the application allows you to generate a rep

Página 356 - 348 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 411Generating Reports2. Press Report.3. In the Save As window, enter a file name.4. From the Save as type li

Página 357

Preparing Your Optical Spectrum Analyzer for a Test34 FTB-5240S/5240BPPerforming User CalibrationTo save a user calibration:1. From the Main Menu, pre

Página 358 - ST Graph Tab

Managing Results412 FTB-5240S/5240BPGenerating ReportsFrom the Save as type list, select txt files (*.txt) to save the report in .txt format (if the f

Página 359 - Optical Spectrum Analyzer 351

Managing ResultsOptical Spectrum Analyzer 413Generating Reports5. Press Save. The report will be added to the Reports folder. You can change the locat

Página 360 - or frequency (nm or Thz):

Managing Results414 FTB-5240S/5240BPGenerating ReportsA sample WDM .txt report is shown below:

Página 361

Managing ResultsOptical Spectrum Analyzer 415Generating ReportsA sample Drift .html report is shown below:

Página 362

A sample Drift .txt report is shown below:

Página 363 - Optical Spectrum Analyzer 355

Managing ResultsOptical Spectrum Analyzer 417Generating ReportsA sample DFB .html report is shown below:

Página 364

Managing Results418 FTB-5240S/5240BPGenerating ReportsA sample Spectral Transmittance .html report is shown below:

Página 365 - Optical Spectrum Analyzer 357

Managing ResultsOptical Spectrum Analyzer 419Generating ReportsA sample EDFA .html report is shown below:

Página 367 - Optical Spectrum Analyzer 359

Optical Spectrum Analyzer 42112 MaintenanceTo help ensure long, trouble-free operation:³ Always inspect fiber-optic connectors before using them and c

Página 368

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 35Performing User Calibration2. Select the Calibration tab.

Página 369 - Optical Spectrum Analyzer 361

Maintenance422 FTB-5240S/5240BPCleaning EUI ConnectorsCleaning EUI ConnectorsRegular cleaning of EUI connectors will help maintain optimum performance

Página 370

MaintenanceOptical Spectrum Analyzer 423Cleaning EUI Connectors5. Repeat steps 3 to 4 with a dry cleaning tip.Note: Make sure you don’t touch the soft

Página 371

Maintenance424 FTB-5240S/5240BPRecalibrating the UnitRecalibrating the UnitManufacturing and service center calibrations are based on the ISO/IEC 1702

Página 372

MaintenanceOptical Spectrum Analyzer 425Recycling and Disposal (Applies to European Union Only)Recycling and Disposal (Applies to European Union Only)

Página 374

Optical Spectrum Analyzer 42713 TroubleshootingViewing Online DocumentationA PDF version of the FTB-5240S/5240BP Optical Spectrum Analyzer user guide

Página 375 - Optical Spectrum Analyzer 367

Troubleshooting428 FTB-5240S/5240BPAboutAbout You can view the product version details and technical support information in this window.To view the in

Página 376 - Managing EDFA Test Results

TroubleshootingOptical Spectrum Analyzer 429About2. Press About.

Página 377 - Optical Spectrum Analyzer 369

Troubleshooting430 FTB-5240S/5240BPTransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transpor

Página 378

Optical Spectrum Analyzer 43114 WarrantyGeneral InformationEXFO Inc. (EXFO) warrants this equipment against defects in material and workmanship for a

Página 379

Preparing Your Optical Spectrum Analyzer for a Test36 FTB-5240S/5240BPPerforming User Calibration3. Press Save Factors, to save the modified user cali

Página 380

Warranty432 FTB-5240S/5240BPLiabilityLiabilityEXFO shall not be liable for damages resulting from the use of the product, nor shall be responsible for

Página 381 - Optical Spectrum Analyzer 373

WarrantyOptical Spectrum Analyzer 433ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products

Página 382 - 374 FTB-5240S/5240BP

Warranty434 FTB-5240S/5240BPService and RepairsService and RepairsEXFO commits to providing product service and repair for five years following the da

Página 383

WarrantyOptical Spectrum Analyzer 435EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your near

Página 385

Optical Spectrum Analyzer 437A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information pres

Página 387

Optical Spectrum Analyzer 439B SCPI Command ReferenceThis appendix presents detailed information on the commands and queries supplied with your FTB-52

Página 388

SCPI Command Reference440 FTB-5240S/5240BPQuick Reference Command TreeQuick Reference Command TreeCommand Parameter(s) P.ABORt 443CALCulate[1..n]BWAT

Página 389 - Optical Spectrum Analyzer 381

SCPI Command ReferenceOptical Spectrum Analyzer 441Quick Reference Command TreeLEVel? 466POWer OFFSet [OUTPut] <Offset> 467[OUTPut]? 468RBW <

Página 390 - 382 FTB-5240S/5240BP

Optical Spectrum Analyzer 375 Setting Up the Instrument in WDM ModeBefore performing a spectral analysis in the WDM mode, you must set up your OSA mod

Página 391

SCPI Command Reference442 FTB-5240S/5240BPQuick Reference Command TreeSTORe CONFiguratio <Filename> 492OVERwrite <Overwrite> 493OVERwrite?

Página 392

SCPI Command ReferenceOptical Spectrum Analyzer 443Product-Specific Commands—DescriptionProduct-Specific Commands—Description:ABORtDescription This co

Página 393

SCPI Command Reference444 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:BWATDescription This command sets the power used to ca

Página 394

SCPI Command ReferenceOptical Spectrum Analyzer 445Product-Specific Commands—Description:CALCulate[1..n]:BWAT?Description This query returns the power

Página 395

SCPI Command Reference446 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel?Description This query returns available infor

Página 396

SCPI Command ReferenceOptical Spectrum Analyzer 447Product-Specific Commands—DescriptionB=Channel Center (always in m) <NR3 NUMERIC RESPONSE DATA&g

Página 397

SCPI Command Reference448 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:ADDDescription Adds a new channel to list of c

Página 398 - 390 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 449Product-Specific Commands—DescriptionCentral wavelenght of the channel being added.Range is depende

Página 399

SCPI Command Reference450 FTB-5240S/5240BPProduct-Specific Commands—Description³ NoiseForOSNR: The program data syntax for the third parameter is defi

Página 400 - 392 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 451Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:COUNt?Description This query returns

Página 401

Setting Up the Instrument in WDM Mode38 FTB-5240S/5240BP³ The most efficient way to setup the instrument is to use one of the favorites configurations

Página 402

SCPI Command Reference452 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:NFOSnrDescription This command sets th

Página 403 - Swapping EDFA Traces

SCPI Command ReferenceOptical Spectrum Analyzer 453Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:NFOSnr?Description This query

Página 404 - Using Zoom Controls

SCPI Command Reference454 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:SIGPowercalcDescription This command s

Página 405

SCPI Command ReferenceOptical Spectrum Analyzer 455Product-Specific Commands—DescriptionPEAKSIGNAL, selects Peak signal power method to calculate sign

Página 406 - Managing Trace Files

SCPI Command Reference456 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:SIGPowercalc?Description This query re

Página 407 - 2. Press New

SCPI Command ReferenceOptical Spectrum Analyzer 457Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault[:STATe]Description This comma

Página 408 - 400 FTB-5240S/5240BP

SCPI Command Reference458 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault[:STATe]?Description This query returns

Página 409 - Opening Trace Files

SCPI Command ReferenceOptical Spectrum Analyzer 459Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:WIDThDescription This command

Página 410

SCPI Command Reference460 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:WIDTh?Description This query returns t

Página 411 - Optical Spectrum Analyzer 403

SCPI Command ReferenceOptical Spectrum Analyzer 461Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DELeteDescription This command delete

Página 412 - ³ Trace identification

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 39Defining PreferencesDefining PreferencesThe preferences window allows you to set gene

Página 413

SCPI Command Reference462 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DELete:ALLDescription This command deletes all

Página 414

SCPI Command ReferenceOptical Spectrum Analyzer 463Product-Specific Commands—Description:CALCulate[1..n][:IMMediate]:AUTODescription This command indi

Página 415 - Optical Spectrum Analyzer 407

SCPI Command Reference464 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n][:IMMediate]:AUTO?Description This query returns the cu

Página 416

SCPI Command ReferenceOptical Spectrum Analyzer 465Product-Specific Commands—Description:CALCulate[1..n]:PEAK[:DETEction]:LEVelDescription This comman

Página 417 - Saving Trace Files

SCPI Command Reference466 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:PEAK[:DETEction]:LEVel?Description This query returns

Página 418 - Generating Reports

SCPI Command ReferenceOptical Spectrum Analyzer 467Product-Specific Commands—Description:CALCulate[1..n]:POWer:OFFSet[:OUTPut]Description This command

Página 419

SCPI Command Reference468 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:POWer:OFFSet[:OUTPut]?Description This querry returns

Página 420

SCPI Command ReferenceOptical Spectrum Analyzer 469Product-Specific Commands—Description:CALCulate[1..n]:RBWDescription This command sets the RBW for

Página 421

SCPI Command Reference470 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:RBW?Description This query gets the value for RBW for

Página 422

SCPI Command ReferenceOptical Spectrum Analyzer 471Product-Specific Commands—Description:CALCulate[1..n]:RBW:INSTrumentDescription This command sets w

Página 423

Setting Up the Instrument in WDM Mode40 FTB-5240S/5240BPDefining Preferences2. Select the General tab.3. Define the general parameters as needed.4. Pr

Página 424

SCPI Command Reference472 FTB-5240S/5240BPProduct-Specific Commands—DescriptionThe <RbwInstrumentState> parameter is a boolean value that determ

Página 425

SCPI Command ReferenceOptical Spectrum Analyzer 473Product-Specific Commands—Description:CALCulate[1..n]:RBW:INSTrument?Description This query returns

Página 426

SCPI Command Reference474 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WAVelength:OFFSet[:OUTPut]Description This command set

Página 427

SCPI Command ReferenceOptical Spectrum Analyzer 475Product-Specific Commands—Description:CALCulate[1..n]:WAVelength:OFFSet[:OUTPut]?Description This q

Página 428

SCPI Command Reference476 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDMDescription When used alone, this command triggers

Página 429 - 12 Maintenance

SCPI Command ReferenceOptical Spectrum Analyzer 477Product-Specific Commands—Description:CALCulate[1..n]:WDM:CHANnel?Description This query returns th

Página 430

SCPI Command Reference478 FTB-5240S/5240BPProduct-Specific Commands—DescriptionE = Noise (always in dBm) <NR3 NUMERIC RESPONSE DATA>F = Noise ty

Página 431

SCPI Command ReferenceOptical Spectrum Analyzer 479Product-Specific Commands—Description:CALCulate[1..n]:WDM:CHANnel:COUNt?Description This query retu

Página 432 - Recalibrating the Unit

SCPI Command Reference480 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:OSNR:AVERage?Description This query returns the av

Página 433 - Recycling and Disposal

SCPI Command ReferenceOptical Spectrum Analyzer 481Product-Specific Commands—Description:CALCulate[1..n]:WDM:OSNR:FLATness?Description This query retu

Página 434

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 41Defining PreferencesTo enter link and location information:1. From the Main Menu, pre

Página 435 - 13 Troubleshooting

SCPI Command Reference482 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:AVERage?Description This query returns th

Página 436 - Troubleshooting

SCPI Command ReferenceOptical Spectrum Analyzer 483Product-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:FLATness?Description This query

Página 437 - 2. Press About

SCPI Command Reference484 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:INTegrated?Description This query returns

Página 438 - Transportation

SCPI Command ReferenceOptical Spectrum Analyzer 485Product-Specific Commands—Description:CALibration:ZERO[:IMMediate]Description This command performs

Página 439 - 14 Warranty

SCPI Command Reference486 FTB-5240S/5240BPProduct-Specific Commands—Description:INITiate[1..n][:IMMediate]Description This command completes one full

Página 440 - Liability

SCPI Command ReferenceOptical Spectrum Analyzer 487Product-Specific Commands—Description:MMEMory[1..n]:LOAD:CONFiguratioDescription This command loads

Página 441

SCPI Command Reference488 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:LOAD:OVERwriteDescription This command selects, whether

Página 442 - Service and Repairs

SCPI Command ReferenceOptical Spectrum Analyzer 489Product-Specific Commands—Description:MMEMory[1..n]:LOAD:OVERwrite?Description This query lets you

Página 443 - Optical Spectrum Analyzer 435

SCPI Command Reference490 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:LOAD:TRACeDescription This command loads a trace.For thi

Página 444

SCPI Command ReferenceOptical Spectrum Analyzer 491Product-Specific Commands—DescriptionSelects Trace index of destination.TRCMAIN, selects Main Trace

Página 445 - A Technical Specifications

ContentsOptical Spectrum Analyzer vA Technical Specifications ... 437B SCPI C

Página 446

Setting Up the Instrument in WDM Mode42 FTB-5240S/5240BPDefining Preferences3. Under System and link information, define the following parameters as n

Página 447 - B SCPI Command Reference

SCPI Command Reference492 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:STORe:CONFiguratioDescription This command saves a confi

Página 448 - Quick Reference Command Tree

SCPI Command ReferenceOptical Spectrum Analyzer 493Product-Specific Commands—Description:MMEMory[1..n]:STORe:OVERwriteDescription This command sets, w

Página 449 - SCPI Command Reference

SCPI Command Reference494 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:STORe:OVERwrite?Description This query lets you know whe

Página 450

SCPI Command ReferenceOptical Spectrum Analyzer 495Product-Specific Commands—Description:MMEMory[1..n]:STORe:TRACeDescription This command saves trace

Página 451 - Optical Spectrum Analyzer 443

SCPI Command Reference496 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:COUNtDescription Sets the number of scans that wil

Página 452 - :CALCulate[1..n]:BWAT

SCPI Command ReferenceOptical Spectrum Analyzer 497Product-Specific Commands—Description:SENSe[1..n]:AVERage:COUNt?Description This query returns numb

Página 453 - :CALCulate[1..n]:BWAT?

SCPI Command Reference498 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:INBand:COUNtDescription This command sets the numb

Página 454 - :CALCulate[1..n]:CHANnel?

SCPI Command ReferenceOptical Spectrum Analyzer 499Product-Specific Commands—Description:SENSe[1..n]:AVERage:INBand:COUNt?Description This query retur

Página 455

SCPI Command Reference500 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:TYPEDescription This command sets the acquisition

Página 456 - :CALCulate[1..n]:CHANnel:ADD

SCPI Command ReferenceOptical Spectrum Analyzer 501Product-Specific Commands—Description:SENSe[1..n]:AVERage:TYPE?Description This query returns the a

Página 457

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 43Defining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Página 458

SCPI Command Reference502 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STARtDescription This command sets the

Página 459 - Optical Spectrum Analyzer 451

SCPI Command ReferenceOptical Spectrum Analyzer 503Product-Specific Commands—DescriptionExample(s) SENS:WAV:RANG:STAR 1520 NMNotes SENS:WAV:RANG:STOP

Página 460 - 452 FTB-5240S/5240BP

SCPI Command Reference504 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STARt?Description This query returns th

Página 461 - Optical Spectrum Analyzer 453

SCPI Command ReferenceOptical Spectrum Analyzer 505Product-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STOPDescription This command s

Página 462 - SIGPowercalc

SCPI Command Reference506 FTB-5240S/5240BPProduct-Specific Commands—DescriptionExample(s) SENS:WAV:RANG:STOP 1570 NMNotes SENS:WAV:RANG:STARt value wi

Página 463

SCPI Command ReferenceOptical Spectrum Analyzer 507Product-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STOP?Description This query re

Página 464 - SIGPowercalc?

SCPI Command Reference508 FTB-5240S/5240BPProduct-Specific Commands—Description:SNUMber?Description This query returns a value indicating the serial n

Página 465 - [:STATe]

SCPI Command ReferenceOptical Spectrum Analyzer 509Product-Specific Commands—Description:STATus?Description This query returns a value indicating the

Página 466 - [:STATe]?

SCPI Command Reference510 FTB-5240S/5240BPProduct-Specific Commands—DescriptionDEFECTIVE, means the module is defective andUNCONFIGURED, means the mod

Página 467 - ³ ChannelWidthUnit:

SCPI Command ReferenceOptical Spectrum Analyzer 511Product-Specific Commands—Description:TRACe[1..n][:DATA]?Description This query returns the points

Página 468 - 460 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode44 FTB-5240S/5240BPDefining Preferences2. Select the Comments tab. 3. Enter your comments for the current trace.

Página 469 - Optical Spectrum Analyzer 461

SCPI Command Reference512 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n][:DATA]:PREamble?Description This query returns a trace’s h

Página 470 - 462 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 513Product-Specific Commands—DescriptionA = Acquisition type <STRING RESPONSE DATA>B = Spectral

Página 471 - Optical Spectrum Analyzer 463

SCPI Command Reference514 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n]:INFormation?Description This query returns all available i

Página 472 - 464 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 515Product-Specific Commands—DescriptionA = Acquisition start date and time <STRING RESPONSE DATA&g

Página 473 - Optical Spectrum Analyzer 465

SCPI Command Reference516 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n]:POINts?Description This query returns the number of points

Página 474 - 466 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 517Product-Specific Commands—Description:UNIT[1..n]:SPECtrumDescription This command sets the spectrum

Página 475 - Parameter(s) Offset:

SCPI Command Reference518 FTB-5240S/5240BPProduct-Specific Commands—Description:UNIT[1..n]:SPECtrum?Description This query returns the current spectru

Página 476 - for global channel

Optical Spectrum Analyzer 519C Formulas Used with Your Optical Spectrum AnalyzerThe following formulas are used in the various tests available with yo

Página 477 - :CALCulate[1..n]:RBW

Formulas Used with Your Optical Spectrum Analyzer520 FTB-5240S/5240BPCentral Wavelength Calculation (Spectral Transmittance)Central Wavelength Calcula

Página 478 - :CALCulate[1..n]:RBW?

Formulas Used with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 521Bandwidth Calculation (Spectral Transmittance)Bandwidth Calculation (Spe

Página 479 - Optical Spectrum Analyzer 471

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 45Defining PreferencesDefining Display ParametersThe application allows you to set disp

Página 481 - Optical Spectrum Analyzer 473

IndexOptical Spectrum Analyzer 523IndexAacquisitionaveraging... 99, 159, 179, 205, 265real-time ... 99, 179, 205, 265

Página 482 - Range: [-99.999 … 99.999] nm

Index524 FTB-5240S/5240BPFFabry-Perot mode spacing... 343fiber ends, cleaning... 15fiber ID ..

Página 483 - [:OUTPut]?

IndexOptical Spectrum Analyzer 525wavelength... 64, 129online user guide... 427op

Página 484 - :CALCulate[1..n]:WDM

Index526 FTB-5240S/5240BPspecifications, product ... 437statusfail ...

Página 485 - :CALCulate[1..n]:WDM:CHANnel?

NOTICE抩⛙CHINESE REGULATION ON RESTRICTION OF HAZARDOUS SUBSTANCES₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩NAMES AND CONTENTS OF THE TOXIC OR HAZARDOUS SUBSTANCES OR ELEMENTS CONT

Página 486

MARKING REQUIREMENTS㪖㽷尐㻑Productℶ❐Environmental protection use period (years)䘾⬒≬㔳∎䞷㦮棟 ( ㄃ )Logo㪖㉦This Exfo product㦻 EXFO ℶ❐10Batterya䟄㻯aa. If applicabl

Página 487 - Optical Spectrum Analyzer 479

www.EXFO.com · [email protected]/N: 1059976CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-

Página 488 - 480 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode46 FTB-5240S/5240BPDefining Preferences2. Select the Display tab.

Página 489 - Optical Spectrum Analyzer 481

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 47Defining Preferences3. Select the spectral unit you want to work with. With the nanom

Página 490 - AVERage?

Setting Up the Instrument in WDM Mode48 FTB-5240S/5240BPDefining Preferences4. Select the label that will appear on the peaks in the graph.

Página 491 - FLATness?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 49Defining PreferencesNote: The channel name and channel number cannot be shown at the

Página 492 - INTegrated?

Setting Up the Instrument in WDM Mode50 FTB-5240S/5240BPDefining Preferences5. Select Show to display the empty channels from the channel list in the

Página 493 - :CALibration:ZERO[:IMMediate]

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 51Defining Preferences7. Select the background color scheme for Graph as desired. 8. Pr

Página 494 - :INITiate[1..n][:IMMediate]

Certification Informationvi FTB-5240S/5240BPCertification InformationF.C.C. InformationElectronic test equipment is exempt from Part 15 compliance (FC

Página 495 - Optical Spectrum Analyzer 487

Setting Up the Instrument in WDM Mode52 FTB-5240S/5240BPDefining PreferencesCustomizing WDM Results TableIt is possible to select which results you wo

Página 496 - :MMEMory[1..n]:LOAD:OVERwrite

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 53Defining Preferences2. Select the WDM Results tab. 3. Select which parameters you wan

Página 497 - Optical Spectrum Analyzer 489

Setting Up the Instrument in WDM Mode54 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the columns will appear i

Página 498 - :MMEMory[1..n]:LOAD:TRACe

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 55Defining PreferencesCustomizing WDM File NameThe application shall provide a way to d

Página 499

Setting Up the Instrument in WDM Mode56 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameters you want to includ

Página 500 - 492 FTB-5240S/5240BP

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 57Defining Preferences4. Press Up or Down to change the order in which the field values

Página 501 - Optical Spectrum Analyzer 493

Setting Up the Instrument in WDM Mode58 FTB-5240S/5240BPSetting Up WDM Analysis ParametersSetting Up WDM Analysis ParametersThis section presents the

Página 502 - 494 FTB-5240S/5240BP

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 59Setting Up WDM Analysis ParametersDefining General SettingsThe general analysis param

Página 503 - :MMEMory[1..n]:STORe:TRACe

Setting Up the Instrument in WDM Mode60 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the General tab.

Página 504 - :SENSe[1..n]:AVERage:COUNt

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 61Setting Up WDM Analysis Parameters3. Under Default channel settings, define the follo

Página 505 - :SENSe[1..n]:AVERage:COUNt?

Certification InformationOptical Spectrum Analyzer vii Application of Council Directive(s): 2006/95/EC - The Low Voltage Directive 2004/108/EC -

Página 506 - 498 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode62 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Channel width (GHz or nm): indicates the limit inside whic

Página 507 - Optical Spectrum Analyzer 499

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 63Setting Up WDM Analysis Parameters³ Noise for OSNR: indicates which calculation metho

Página 508 - :SENSe[1..n]:AVERage:TYPE

Setting Up the Instrument in WDM Mode64 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Peak detection level (dBm): indicates the minimum power le

Página 509 - :SENSe[1..n]:AVERage:TYPE?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 65Setting Up WDM Analysis Parameters5. If you want to apply the above settings to the a

Página 510 - 502 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode66 FTB-5240S/5240BPSetting Up WDM Analysis ParametersYou can set your pass/fail threshold limits in different way

Página 511 - Optical Spectrum Analyzer 503

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 67Setting Up WDM Analysis ParametersDefining Global ThresholdsGlobal thresholds will be

Página 512 - 504 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode68 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo define global thresholds:1. From the Main Menu, press Ana

Página 513 - Optical Spectrum Analyzer 505

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 69Setting Up WDM Analysis Parameters3. Select the Activate all thresholds option, to ma

Página 514 - 506 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode70 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Empty channel count: Select this option to obtain the numb

Página 515 - Optical Spectrum Analyzer 507

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 71Setting Up WDM Analysis ParametersDefining Default ThresholdsDefault thresholds will

Página 517 - :STATus?

Setting Up the Instrument in WDM Mode72 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Default Thresholds tab. 3. Enter values in the

Página 518

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 73Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Página 519 - :TRACe[1..n][:DATA]?

Setting Up the Instrument in WDM Mode74 FTB-5240S/5240BPSetting Up WDM Analysis ParametersManaging ChannelsTesting DWDM systems involves characterizin

Página 520 - :TRACe[1..n][:DATA]:PREamble?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 75Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. By default, the chan

Página 521

Setting Up the Instrument in WDM Mode76 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Stop range (nm or Thz): indicates the ending range of the

Página 522 - :TRACe[1..n]:INFormation?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 77Setting Up WDM Analysis Parameters6. If you want to apply the above settings to the a

Página 523

Setting Up the Instrument in WDM Mode78 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo edit the parameters of a specific channel:1. From the Mai

Página 524 - :TRACe[1..n]:POINts?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 79Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. Select the channel o

Página 525 - :UNIT[1..n]:SPECtrum

Setting Up the Instrument in WDM Mode80 FTB-5240S/5240BPSetting Up WDM Analysis Parameters4. Press Edit Selection.

Página 526 - :UNIT[1..n]:SPECtrum?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 81Setting Up WDM Analysis Parameters5. Modify the settings as needed. For more informat

Página 527 - C Formulas Used with Your

Optical Spectrum Analyzer 11 Introducing the FTB-5240S/5240BP Optical Spectrum AnalyzerThe FTB-5240S/5240BP Optical Spectrum Analyzer (OSA) is designe

Página 528 - Transmittance)

Setting Up the Instrument in WDM Mode82 FTB-5240S/5240BPSetting Up WDM Analysis Parameters6. Press OK to return to the Channels tab, which now contain

Página 529

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 83Setting Up WDM Analysis ParametersTo add current peaks:Note: You can add current peak

Página 530

Setting Up the Instrument in WDM Mode84 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Channels tab. 3. Press Import from Trace. All

Página 531 - Optical Spectrum Analyzer 523

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 85Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Página 532

Setting Up the Instrument in WDM Mode86 FTB-5240S/5240BPSetting Up WDM Analysis ParametersManaging FavoritesFavorites are configurations files that co

Página 533 - Optical Spectrum Analyzer 525

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 87Setting Up WDM Analysis Parameters2. Select the Favorites tab.

Página 534

Setting Up the Instrument in WDM Mode88 FTB-5240S/5240BPSetting Up WDM Analysis Parameters3. To apply the settings from a favorite file to the current

Página 535 - ₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 89Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Página 536 - MARKING REQUIREMENTS

Setting Up the Instrument in WDM Mode90 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo save a test configuration:1. From the Main Menu, press An

Página 537

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 91Setting Up WDM Analysis Parameters2. Select the Favorites tab.3. To save an analysis