Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manual do Utilizador Página 431

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 659
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 430
Referencia de instrucciones SCPI
OSA 425
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
CENTer:FREQuency
Description This command sets the nominal center
frequency of the selected WDM analysis channel.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:CHANnel:CENTer:FREQ
uency<wsp><Center[<wsp>HZ]>|MAXimum
|MINimum|DEFault
Parameter(s) Center:
The program data syntax for <Center> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is HZ. The <Center>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Vista de página 430
1 2 ... 426 427 428 429 430 431 432 433 434 435 436 ... 658 659

Comentários a estes Manuais

Sem comentários