SCPI-Befehlsreferenz
388 Serie FTB-7000
Product-Specific Commands—Description
:CONFigure[1..n]:ACQuisition:RANGe?
Description This query returns the current range setting.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ACQuisition:RANGe?
Parameter(s) None
Response Syntax <Range>
Response(s) Range:
The response data syntax for <Range> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the range, in meters.
Example(s) CONF:ACQ 1310 NM,1250 M,10 NS
CONF:ACQ:RANG? Returns 1.25E+3
See Also CONFigure[1..n]:ACQuisition:WAVelength:LIST?
CONFigure[1..n]:ACQuisition:RANGe:LIST?