Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual do Utilizador Página 463

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 688
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 462
Справка по командам SCPI
OSA 457
Product-Specific Commands—Description
Response(s) Center:
The response data syntax for <Center> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Center> response corresponds to either the
current or the MINimum/MAXimum channel
center wavelength in meter.
Example(s) CALC:CHAN:DEF "CWDM_7",1450.0 NM
CALC:CHAN:SEL "CWDM_7"
CALC:CHAN:CENT:WAV? Returns
1.45000E-006
See Also :CALCulate[1..n][:WDM]:CHANnel[:DEFine]
:CALCulate[1..n][:WDM]:CHANnel:CENTer[:WAV
elength]
:CALCulate[1..n][:WDM]:CHANnel:CENTer:FRE
Quency?
:CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:CHANnel:
CENTer[:WAVelength]?
Vista de página 462
1 2 ... 458 459 460 461 462 463 464 465 466 467 468 ... 687 688

Comentários a estes Manuais

Sem comentários