Справка по командам SCPI
590 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:TYPE
Description This command selects the acquired trace
averaging type.
At *RST, averaging is set to SCALar.
Syntax :SENSe[1..n]:AVERage:TYPE<wsp>SCALar|Pol
arizationMinMaxHold
Parameter(s) Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
SCALar|PolarizationMinMaxHold.
The parameter corresponds to the newly selected
trace averaging type.
SCALar: selects scalar averaging type.
PolarizationMinMaxHold: selects averaging type
for InBand noise measurement.
Example(s) SENS:AVER:TYPE SCAL
SENS:AVER:TYPE? Returns SCALAR