Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual do Utilizador Página 560

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 688
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 559
Справка по командам SCPI
554 FTB-5240S/S-P/BP
Product-Specific Commands—Description
The <Auto> parameter corresponds to the new
state of auto resolution bandwidth for OSNR
calculation.
0 or OFF: custom resolution bandwidth value is
used.
1 or ON: instruments resolution bandwidth is
used.
Example(s) CALC:WDM:OSNR:BAND:RES:AUTO ON
CALC:WDM:OSNR:BAND:RES:AUTO? Returns
1 (instrument's RBW enabled)
CALC:WDM:OSNR:BAND:RES 0.100 NM
CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES:AUTO? Returns
0 (RBW 0.100 nm enabled)
See Also :CALCulate[1..n][:WDM]:BWIDth[1|2]|BANDwidth[
1|2]:RelativeLEVel
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]:AUTO?
:CALCulate[1..n][:WDM]:THReshold
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]:AUTO
Vista de página 559
1 2 ... 555 556 557 558 559 560 561 562 563 564 565 ... 687 688

Comentários a estes Manuais

Sem comentários