Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual do Utilizador Página 529

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 528
SCPI Command Reference
OSA 523
Product-Specific Commands—Description
With <Continuous> set to OFF, the trigger
system remain in idle state until <Continuous>
is set to ON or INITiate:IMMediate command is
received. With <Continuous> set to ON, the
trigger system leave idle state and continue
cycling until <Continuous> set to OFF or ABORt
command is received.
When <Continuous> is set to OFF, the current
trigger cycle is completed before returning to idle
state: current acquisition continue until finished.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:AUTO ON
CALC:WDM:CHAN:AUTO:NOIS:AUTO OFF
CALC:WDM:CHAN:AUTO:NOIS:TYPE IEC
CALC:WDM:STATe ON
TRACe:FEED:CONTrol "TRC1", ALW
SENS:AVER:STAT OFF
:INITiate:CONTinuous
Vista de página 528
1 2 ... 524 525 526 527 528 529 530 531 532 533 534 ... 637 638

Comentários a estes Manuais

Sem comentários