SCPI Command Reference
OSA 559
Product-Specific Commands—Description
:SENSe[1..n][:WAVelength]:OFFSet
Description This command sets the wavelength offset that is
added to every point measured by the
instrument.
At *RST, this value is set to 0.0 nm.
Syntax :SENSe[1..n][:WAVelength]:OFFSet<wsp><Offs
et[<wsp>M]>|MAXimum|MINimum|DEFault
Parameter(s) Offset:
The program data syntax for <Offset> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is M. The <Offset> special forms
MINimum, MAXimum and DEFault are accepted
on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.