Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual do Utilizador Página 530

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 529
SCPI Command Reference
524 FTB-5240S/S-P/BP
Product-Specific Commands—Description
SENS:WAV:STAR 1525.000 NM
SENS:WAV:STOP 1570.000 NM
TRIG:SEQ:SOUR IMM
STAT? Poll until returned state is READY
INIT:CONT ON
INIT:CONT? Returns 1 (trigger system
continuously initiated)
...
INIT:CONT OFF
STAT:OPER:BIT8:COND? Poll until returned state
is 0
Notes The trigger system leaves IDLE state to perform
acquisition only if the instrument is in READY
status.
Trace averaging is not supported by the trigger
system when continuously initiated.
Continuous acquisition does not support InBand
noise analysis: acquired trace is always analysed
using IEC noise measurement.
See Also :ABORt
:CALCulate[1..n][:WDM]:STATe
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:SENSe[1..n]:AVERage:STATe
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?
:TRACe:FEED:CONTrol
:TRIGger[1..n][:SEQuence]:SOURce
:INITiate:CONTinuous
Vista de página 529
1 2 ... 525 526 527 528 529 530 531 532 533 534 535 ... 637 638

Comentários a estes Manuais

Sem comentários