Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual do Utilizador Página 612

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 611
SCPI Command Reference
606 FTB-5240S/S-P/BP
Examples on Using the SCPI Commands
Modifying Global Analysis Parameters (WDM)
Command Sequence Comments
CALC:WDM:THR -45.00 DBM Set channel peak detection level.
CALC:WDM:OSNR:BAND:RES:AUTO
OFF
Select between the instrument's native or
custom resolution bandwidth for OSNR
computing.
CALC:WDM:OSNR:BAND:RES 0.100
NM
Set the custom resolution bandwidth for OSNR.
CALC:WDM:BAND2:RLEV 20.0 DB Set the user defined bandwidth position for all
channels.
Vista de página 611
1 2 ... 607 608 609 610 611 612 613 614 615 616 617 ... 637 638

Comentários a estes Manuais

Sem comentários