Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manual do Utilizador Página 546

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 545
SCPI Command Reference
540 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:COUNt?
Description This query returns a value indicating either the
current or the minimum/maximum number of
acquired traces to combine for averaging.
At *RST, averaging count is set to 8.
Syntax :SENSe[1..n]:AVERage:COUNt?[<wsp>MAXimu
m|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Count>
Vista de página 545
1 2 ... 541 542 543 544 545 546 547 548 549 550 551 ... 637 638

Comentários a estes Manuais

Sem comentários