Exfo IQS-12004B DWDM Passive Component Test System for Manual do Utilizador Página 251

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 250
SCPI Commands Reference
DWDM Passive Component Test System 239
Specific Commands
CONFiguration:AutoRangeFiltering
Description This command sets the filtering mode.
Syntax CONF:ARF<wsp><filtering>
Parameter The <filtering> parameter is a boolean value indicating if it is in
filtering mode:
“1” - Sets the system to filtering mode
“0” - Does not set the system to filtering mode
Response None
Example CONF:ARF 1
Note It is highly recommended to set the system to filtering mode.
CONFiguration:AutoRangeFiltering?
Description The response to this query indicates the filtering mode.
Syntax CONF:ARF?
Parameter None
Response A value representing the selected filtering mode:
“1” - The system is in filtering mode
“0” - The system is not in filtering mode
Example CONF:ARF?
Vista de página 250
1 2 ... 246 247 248 249 250 251 252 253 254 255 256 ... 436 437

Comentários a estes Manuais

Sem comentários