COM Objects Reference
356 IQS-12004B
Analysis Interface—Methods
Analysis
SetIlTrace, SetOrlTrace, SetPdlTrace
Description Each of these methods allows to re-analyze a previously
acquired trace.
Syntax object.SetIlTrace (Trace)
object.SetOrlTrace (Trace)
object.SetPdlTrace (Trace)
Parameters Trace: Required. A Variant value containing a two-dimensional
array representing pairs of X,Y coordinates of the trace to
re-analyze.
Response None
Note The SetIlTrace, SetOrlTrace and SetPdlTrace methods let you
re-analyze any acquired trace at a later time. This is mainly
useful when you want to compare the results before and after
applying drift traces for environmental type tests.
Once the trace of interest is set, you must start the DoAnalysis
method before retrieving results.
Use this method in conjunction with the following methods:
AddRefExtZeroTrace, AddRefExtTrace, GetDriftTrace,
GetCorrectedIlTrace.