SCPI Commands Reference
DWDM Passive Component Test System 259
Specific Commands
READ:CHANnel?
Description The response to this query indicates the active channel.
Syntax READ:CHAN?
Parameter None
Response Active channel number in the “999” format.
Example READ:CHAN?
READ:CHannelLOcalization?
Description The response to this query indicates the localization
information about the specified channels.
Syntax READ:CHLO?
Parameter None
Response A string response in the “Slot Position,Power Meter Channel”
format. For example:
- Slot position: (1-3)
- Power meter channel: 2 (range: 1-4)
Example READ:CHLO?