COM Objects Reference
278 IQS-12004B
Measures Interface—Methods
Measures
CalWr
Description This command performs a wavelength response calibration on
the power meter channel.
ChannelToCalibrate represents the number of channels to
calibrate, as indicated on the channel localization list.
Syntax object.CalWr(ChannelToCalibrate, FireEvent)
Parameters Byval Long ChannelToCalibrate, Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrInProgress, ErrChannelNumber, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more
information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages on page 208.
See also ChannelLocalisation (see page 291)
Measures
RefIl
Description This command performs an IL reference measurement.
Syntax object.RefIl(FireEvent)
Parameters Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrCalWrBefore, InProgress, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more
information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages on page 208.